Welch UHPLC-MS/MS Solution for PFAS Analysis in Food

Welch UHPLC-MS/MS Solution for PFAS Analysis in Food

Introduction

Per- and polyfluoroalkyl substances (PFAS) are a large group of synthetic fluorinated organic compounds characterized by strong carbon–fluorine (C–F) bonds. Their excellent water and oil repellency, thermal stability, and chemical resistance have made them widely used in industrial and consumer applications, including textile treatments, food contact materials, firefighting foams, paper manufacturing, electroplating, and coatings.

However, many PFAS are highly persistent in the environment, and some can accumulate in living organisms. Human exposure may occur through drinking water, food, air, and direct contact with contaminated materials. Growing concerns over their potential health risks have prompted regulatory authorities worldwide to introduce increasingly stringent controls on PFAS in environmental and food-related matrices.

Analytical Challenges and Solution

PFAS encompass a diverse range of compounds, from legacy perfluorocarboxylic acids (PFCAs) and perfluorosulfonic acids (PFSAs) to fluorotelomer sulfonic acids and emerging alternatives. Their diverse chemical properties, combined with the complexity of food matrices, place high demands on analytical methods in terms of chromatographic separation, detection sensitivity, and interference control.

To address these challenges, Welch Materials has developed a UHPLC-MS/MS solution for the simultaneous determination of 25 PFAS compounds in food. Featuring a Welch Xtimate UHPLC C18 column coupled with tandem mass spectrometry, the solution enables efficient separation and sensitive, selective detection of a broad range of PFAS.

Class
Analytes
PFCAs PFBA, PFPeA, PFHxA, PFHpA, PFOA, PFNA, PFDA, PFUnDA, PFDoDA
PFSAs PFBS, PFPeS, PFHpS, PFHxS, PFOS
FTSAs 4:2FTSA, 6:2FTSA, 8:2FTSA
Novel PFAS HFPO-DA, PFMPA, PFMBA, PFEESA, NFDHA, DONA, 9Cl-PF3ONS, 11Cl-PF3OUdS

Chromatographic and Mass Spectrometry Conditions

UHPLC conditions:

  • Column: Xtimate UHPLC C18 2.1×50 mm, 1.8 µm
  • Mobile phase: A: 5 mmol/L ammonium acetate; B: Methanol
  • Flow rate: 0.4 mL/min
  • Column temperature: 30 °C
  • Injection volume: 10 µL
Time (min) Phase A (%) Phase B (%)
0.00 85 15
1.00 85 15
1.50 45 55
5.5 30 70
7.0 20 80
12.0 0 100
14.4 0 100
20.0 85 15

MS conditions:

  • Ion source & scanning method: ESI-
  • Detection method: MRM
  • Ion spray voltage: -4500V; temperature: 200 °C
  • Gas source: curtain 10 psi, atomization 20 psi, auxiliary 10 psi
  • Instrument: Waters AB Sciex TRIPLE QUAD 6500
Analyte Precursor Ion (m/z) Product Ion (m/z) DP (V) CE (V)
PFBA 213 169* -14 -12
213 -18 -5
PFPeA 263 219* -14 -10
263 -5
PFHxA 313 269* -10 -13
119 -26
PFHpA 363 319* -16 -13
169 -22
PFOA 413 369* -24 -14
169 -23
PFNA 463 419* -95 -14
219 -22
PFDA 513 469* -18 -15
268.9 -25
PFUnDA 563 519* -8 -17
318.9 -24
PFDoDA 613 569* -26 -20
318.9 -26
PFBS 299 80* -16 -60
99 -37
PFPeS 349 80* -8 -78
99 -36
PFHpS 449 80* -14 -87
99 -98
4:2FTSA 327 307* -18 -26
81 -72
6:2FTSA 427 407* -16 -32
80.9 -78
8:2FTSA 527 507* -21 -35
81 -36 -110
HFPO-DA 285 169* -65 -10
185 -19
PFMPA 229 85* -16 -18
229 -5
PFMBA 279 85* -9 -12
235 -10
PFEESA 315 135* -16 -26
82.8 -25
NFDHA 201 85* -88 -19
135 -14
DONA 377 251* -20 -18
84.8 -41
PFHxS 399 80* -9 -86
99 -76
PFOS 499 80* -36 -112
99 -94
9Cl-PF3ONS 531 351* -5 -39
531 -22
11Cl-PF3OUdS 631 451* -18 -41
631 -12
Note: * denotes the quantitation ion.

Sample Preparation

Extraction

  • Chicken breast: Weigh 1 g of defrosted and homogenized sample (accurate to 0.001 g, except for liquid milk samples) and place in a 50 mL polypropylene centrifuge tube, add 10 μL of the 0.1 μg/mL mixed standard working solution, and mix well. Add 10 mL of the 50 mmol/L NaOH-in-Methanol solution, rotate the tube in a rotater for 1 hour, and centrifuge at 8,000 r/min for 15 mins. Take the supernatant and nitrogen evaporate to near dryness (less than 500 µL) in a 40 °C water bath. Add 8 mL of water and mix well.
  • Milk: Weight 2 g of the sample (accurate to 0.001 g) and place in a 15 mL polypropylene centrifuge tube. Add 6 mL of acetonitrile, sonicate for 20 mins, and centrifuge at 8,000 r/min for 10 mins. Collect the supernatant in another 15 mL polypropylene centrifuge tube. Add another 6 mL of acetonitrile to the residue and repeat the extraction. Combine the extract, nitrogen evaporate to near dryness (less than 500 µL) in a 40 °C water bath, add 8 mL of water, and mix well.

Cleanup

  • Cartridge: Welchrom PWAX, 500 mg/6 mL.
  • Conditioning: 10 mL methanol, followed by 10 mL of the 0.1 mol/L phosphate buffer (pH 7.0); flow rate approx. 1 mL/min. After the liquid has drained, close the valve and add 3 mL of the 0.1 mol/L phosphate buffer (pH 7.0).
  • Loading: Load the entire solution at approx. 1 mL/min. Discard the effluent.
  • Rinsing: 10 mL of 1 g/L ammonium acetate. Discard the effluent. Then dry.
  • Elution: 10 mL of 2% ammonium hydroxide-methanol. Collect the eluate.
  • Concentration and Reconstitution: Nitrogen evaporate the collected eluate in a water bath at 55–60 °C to dryness. Reconstitute with accurate 1 mL of methanol-water (80:20) solution and dissolve thoroughly. Filter through a 0.22 µm membrane. Proceed to LC-MS/MS analysis.

Analytical Results and Chromatograms

Under a spiking level of 1 µg/kg, the average recovery rates and RSD of the 25 PFAS analytes in chicken breast and milk are listed below.

Analyte
Chicken Breast
Milk
Avg. Recovery RSD (N≥3) Avg. Recovery RSD (N≥3)
PFBA 67.47% 9.23% 87.35% 3.14%
PFPeA 89.33% 6.87% 102.94% 6.69%
PFHxA 80.65% 6.87% 91.78% 1.52%
PFHpA 75.50% 1.45% 88.56% 6.34%
PFOA 85.93% 4.57% 95.69% 4.30%
PFNA 85.96% 3.24% 91.95% 6.82%
PFDA 76.33% 3.56% 95.90% 5.78%
PFUnDA 82.13% 9.87% 66.67% 6.48%
PFDoDA 74.69% 4.99% 66.16% 9.34%
PFBS 89.43% 4.05% 92.79% 8.79%
PFPeS 81.42% 2.58% 109.29% 9.65%
PFHpS 68.68% 2.80% 92.61% 6.23%
4:2FTSA 75.75% 3.80% 90.61% 5.53%
6:2FTSA 78.50% 7.34% 109.69% 4.13%
8:2FTSA 95.47% 2.77% 98.25% 5.97%
HFPO-DA 83.88% 8.45% 98.69% 9.71%
PFMPA 82.72% 3.71% 89.91% 9.01%
PFMBA 80.29% 3.76% 90.64% 5.76%
PFEESA 79.69% 0.28% 111.02% 9.51%
NFDHA 75.29% 6.69% 89.60% 4.73%
DONA 93.56% 9.35% 93.05% 6.19%
PFHxS 81.60% 9.90% 102.13% 3.68%
PFOS 75.71% 5.35% 91.83% 9.08%
9Cl-PF3ONS 82.35% 2.76% 78.41% 9.79%
11Cl-PF3OUdS 74.27% 8.56% 72.27% 7.35%
Mixed Standard Solution Total Ion Chromatogram
Total ion chromatogram of the mixed standard solution (1 µg/L)
Mixed Standard Solution Extracted Ion Chromatogram
Extracted ion chromatogram of the mixed standard solution (1 µg/L)
Chicken Breast Spiked Sample Chromatogram
Chromatogram of spiked chicken breast sample (1 µg/kg)
Milk Spiked Sample Chromatogram
Chromatogram of spiked milk sample (1 µg/kg)

Method Advantages

By combining broad compound coverage with high-efficiency chromatographic separation, the method provides a practical approach to PFAS screening and quantitative analysis in food samples. With a total analytical cycle of 20 minutes, the solution is designed to balance separation performance and laboratory throughput. Its gradient program also incorporates column washing and re-equilibration steps to help minimize carryover and potential cross-contamination between injections.

Ordering Information