Introduction
Per- and polyfluoroalkyl substances (PFAS) are a large group of synthetic fluorinated organic compounds characterized by strong carbon–fluorine (C–F) bonds. Their excellent water and oil repellency, thermal stability, and chemical resistance have made them widely used in industrial and consumer applications, including textile treatments, food contact materials, firefighting foams, paper manufacturing, electroplating, and coatings.
However, many PFAS are highly persistent in the environment, and some can accumulate in living organisms. Human exposure may occur through drinking water, food, air, and direct contact with contaminated materials. Growing concerns over their potential health risks have prompted regulatory authorities worldwide to introduce increasingly stringent controls on PFAS in environmental and food-related matrices.
Analytical Challenges and Solution
PFAS encompass a diverse range of compounds, from legacy perfluorocarboxylic acids (PFCAs) and perfluorosulfonic acids (PFSAs) to fluorotelomer sulfonic acids and emerging alternatives. Their diverse chemical properties, combined with the complexity of food matrices, place high demands on analytical methods in terms of chromatographic separation, detection sensitivity, and interference control.
To address these challenges, Welch Materials has developed a UHPLC-MS/MS solution for the simultaneous determination of 25 PFAS compounds in food. Featuring a Welch Xtimate UHPLC C18 column coupled with tandem mass spectrometry, the solution enables efficient separation and sensitive, selective detection of a broad range of PFAS.
Class |
Analytes |
|---|---|
| PFCAs | PFBA, PFPeA, PFHxA, PFHpA, PFOA, PFNA, PFDA, PFUnDA, PFDoDA |
| PFSAs | PFBS, PFPeS, PFHpS, PFHxS, PFOS |
| FTSAs | 4:2FTSA, 6:2FTSA, 8:2FTSA |
| Novel PFAS | HFPO-DA, PFMPA, PFMBA, PFEESA, NFDHA, DONA, 9Cl-PF3ONS, 11Cl-PF3OUdS |
Chromatographic and Mass Spectrometry Conditions
UHPLC conditions:
- Column: Xtimate UHPLC C18 2.1×50 mm, 1.8 µm
- Mobile phase: A: 5 mmol/L ammonium acetate; B: Methanol
- Flow rate: 0.4 mL/min
- Column temperature: 30 °C
- Injection volume: 10 µL
| Time (min) | Phase A (%) | Phase B (%) |
|---|---|---|
| 0.00 | 85 | 15 |
| 1.00 | 85 | 15 |
| 1.50 | 45 | 55 |
| 5.5 | 30 | 70 |
| 7.0 | 20 | 80 |
| 12.0 | 0 | 100 |
| 14.4 | 0 | 100 |
| 20.0 | 85 | 15 |
MS conditions:
- Ion source & scanning method: ESI-
- Detection method: MRM
- Ion spray voltage: -4500V; temperature: 200 °C
- Gas source: curtain 10 psi, atomization 20 psi, auxiliary 10 psi
- Instrument: Waters AB Sciex TRIPLE QUAD 6500
| Analyte | Precursor Ion (m/z) | Product Ion (m/z) | DP (V) | CE (V) |
|---|---|---|---|---|
| PFBA | 213 | 169* | -14 | -12 |
| 213 | -18 | -5 | ||
| PFPeA | 263 | 219* | -14 | -10 |
| 263 | -5 | |||
| PFHxA | 313 | 269* | -10 | -13 |
| 119 | -26 | |||
| PFHpA | 363 | 319* | -16 | -13 |
| 169 | -22 | |||
| PFOA | 413 | 369* | -24 | -14 |
| 169 | -23 | |||
| PFNA | 463 | 419* | -95 | -14 |
| 219 | -22 | |||
| PFDA | 513 | 469* | -18 | -15 |
| 268.9 | -25 | |||
| PFUnDA | 563 | 519* | -8 | -17 |
| 318.9 | -24 | |||
| PFDoDA | 613 | 569* | -26 | -20 |
| 318.9 | -26 | |||
| PFBS | 299 | 80* | -16 | -60 |
| 99 | -37 | |||
| PFPeS | 349 | 80* | -8 | -78 |
| 99 | -36 | |||
| PFHpS | 449 | 80* | -14 | -87 |
| 99 | -98 | |||
| 4:2FTSA | 327 | 307* | -18 | -26 |
| 81 | -72 | |||
| 6:2FTSA | 427 | 407* | -16 | -32 |
| 80.9 | -78 | |||
| 8:2FTSA | 527 | 507* | -21 | -35 |
| 81 | -36 | -110 | ||
| HFPO-DA | 285 | 169* | -65 | -10 |
| 185 | -19 | |||
| PFMPA | 229 | 85* | -16 | -18 |
| 229 | -5 | |||
| PFMBA | 279 | 85* | -9 | -12 |
| 235 | -10 | |||
| PFEESA | 315 | 135* | -16 | -26 |
| 82.8 | -25 | |||
| NFDHA | 201 | 85* | -88 | -19 |
| 135 | -14 | |||
| DONA | 377 | 251* | -20 | -18 |
| 84.8 | -41 | |||
| PFHxS | 399 | 80* | -9 | -86 |
| 99 | -76 | |||
| PFOS | 499 | 80* | -36 | -112 |
| 99 | -94 | |||
| 9Cl-PF3ONS | 531 | 351* | -5 | -39 |
| 531 | -22 | |||
| 11Cl-PF3OUdS | 631 | 451* | -18 | -41 |
| 631 | -12 | |||
| Note: * denotes the quantitation ion. | ||||
Sample Preparation
Extraction
- Chicken breast: Weigh 1 g of defrosted and homogenized sample (accurate to 0.001 g, except for liquid milk samples) and place in a 50 mL polypropylene centrifuge tube, add 10 μL of the 0.1 μg/mL mixed standard working solution, and mix well. Add 10 mL of the 50 mmol/L NaOH-in-Methanol solution, rotate the tube in a rotater for 1 hour, and centrifuge at 8,000 r/min for 15 mins. Take the supernatant and nitrogen evaporate to near dryness (less than 500 µL) in a 40 °C water bath. Add 8 mL of water and mix well.
- Milk: Weight 2 g of the sample (accurate to 0.001 g) and place in a 15 mL polypropylene centrifuge tube. Add 6 mL of acetonitrile, sonicate for 20 mins, and centrifuge at 8,000 r/min for 10 mins. Collect the supernatant in another 15 mL polypropylene centrifuge tube. Add another 6 mL of acetonitrile to the residue and repeat the extraction. Combine the extract, nitrogen evaporate to near dryness (less than 500 µL) in a 40 °C water bath, add 8 mL of water, and mix well.
Cleanup
- Cartridge: Welchrom PWAX, 500 mg/6 mL.
- Conditioning: 10 mL methanol, followed by 10 mL of the 0.1 mol/L phosphate buffer (pH 7.0); flow rate approx. 1 mL/min. After the liquid has drained, close the valve and add 3 mL of the 0.1 mol/L phosphate buffer (pH 7.0).
- Loading: Load the entire solution at approx. 1 mL/min. Discard the effluent.
- Rinsing: 10 mL of 1 g/L ammonium acetate. Discard the effluent. Then dry.
- Elution: 10 mL of 2% ammonium hydroxide-methanol. Collect the eluate.
- Concentration and Reconstitution: Nitrogen evaporate the collected eluate in a water bath at 55–60 °C to dryness. Reconstitute with accurate 1 mL of methanol-water (80:20) solution and dissolve thoroughly. Filter through a 0.22 µm membrane. Proceed to LC-MS/MS analysis.
Analytical Results and Chromatograms
Under a spiking level of 1 µg/kg, the average recovery rates and RSD of the 25 PFAS analytes in chicken breast and milk are listed below.
Analyte |
Chicken Breast |
Milk |
||
|---|---|---|---|---|
| Avg. Recovery | RSD (N≥3) | Avg. Recovery | RSD (N≥3) | |
| PFBA | 67.47% | 9.23% | 87.35% | 3.14% |
| PFPeA | 89.33% | 6.87% | 102.94% | 6.69% |
| PFHxA | 80.65% | 6.87% | 91.78% | 1.52% |
| PFHpA | 75.50% | 1.45% | 88.56% | 6.34% |
| PFOA | 85.93% | 4.57% | 95.69% | 4.30% |
| PFNA | 85.96% | 3.24% | 91.95% | 6.82% |
| PFDA | 76.33% | 3.56% | 95.90% | 5.78% |
| PFUnDA | 82.13% | 9.87% | 66.67% | 6.48% |
| PFDoDA | 74.69% | 4.99% | 66.16% | 9.34% |
| PFBS | 89.43% | 4.05% | 92.79% | 8.79% |
| PFPeS | 81.42% | 2.58% | 109.29% | 9.65% |
| PFHpS | 68.68% | 2.80% | 92.61% | 6.23% |
| 4:2FTSA | 75.75% | 3.80% | 90.61% | 5.53% |
| 6:2FTSA | 78.50% | 7.34% | 109.69% | 4.13% |
| 8:2FTSA | 95.47% | 2.77% | 98.25% | 5.97% |
| HFPO-DA | 83.88% | 8.45% | 98.69% | 9.71% |
| PFMPA | 82.72% | 3.71% | 89.91% | 9.01% |
| PFMBA | 80.29% | 3.76% | 90.64% | 5.76% |
| PFEESA | 79.69% | 0.28% | 111.02% | 9.51% |
| NFDHA | 75.29% | 6.69% | 89.60% | 4.73% |
| DONA | 93.56% | 9.35% | 93.05% | 6.19% |
| PFHxS | 81.60% | 9.90% | 102.13% | 3.68% |
| PFOS | 75.71% | 5.35% | 91.83% | 9.08% |
| 9Cl-PF3ONS | 82.35% | 2.76% | 78.41% | 9.79% |
| 11Cl-PF3OUdS | 74.27% | 8.56% | 72.27% | 7.35% |
Method Advantages
By combining broad compound coverage with high-efficiency chromatographic separation, the method provides a practical approach to PFAS screening and quantitative analysis in food samples. With a total analytical cycle of 20 minutes, the solution is designed to balance separation performance and laboratory throughput. Its gradient program also incorporates column washing and re-equilibration steps to help minimize carryover and potential cross-contamination between injections.