Introduction
Antibody–drug conjugates (ADCs) play an important role in driving the advance of targeted cancer therapies. As their use expands, the need for efficient, high-precision characterization methods to guarantee safety and efficacy has never been greater.
Welch Materials’ newly launched Advanchrom Bio HIC-Butyl column, with its outstanding separation performance, provides precise, reliable, and regulatory-ready chromatographic analysis of a critical ADC quality attribute — the drug-to-antibody ratio (DAR) — supporting the accelerated translation of innovative ADC candidates into the clinic and market.

Product Information
Advanchrom Bio HIC-Butyl is a high-performance column designed for protein separation based on hydrophobic interaction chromatography (HIC). It utilizes an advanced macroporous silica matrix combined with a unique surface bonding technique.
By precisely optimizing the thickness of the hydrophilic layer and the butyl bonding density on the silica surface, it is well-suited for the separation of antibodies, antibody-drug conjugates (ADCs), and other protein-based biopharmaceutical products.
Compared to Advanchrom HIC-Butyl, Bio HIC-Butyl has a weaker retention.

Product Features
- Unique chemical design offering excellent selectivity for ADC molecules
- Ultra-high purity large-pore silica microsphere matrix for superior column efficiency
- Low non-specific adsorption ensuring high recovery rates
- Excellent pressure tolerance and durability
- Outstanding batch-to-batch reproducibility
Product Specifications
Column | Advanchrom Bio HIC-Butyl |
Matrix | Ultra-high purity large-pore silica |
Particle Size | 5 µm |
Pore Size | 1000 Å |
Column Material | Stainless Steel |
Dimensions | 4.6×100 mm |
pH Range | 2.0-8.0 |
Max Temperature | 60 °C |
Max Pressure Tolerance | 6000 psi |

Application Examples
1. DAR Analysis of ADC Sample
Chromatographic conditions:
- Column: Advanchrom Bio HIC-Butyl (4.6×100 mm, 5 µm)
- Mobile Phases:
A: 50 mM PB solution + 2.0 M ammonium sulfate, pH=7.0
B: 50 mM PB solution, pH=7.0 : isopropanol = 75:25 - Gradient Program:
Time (min) A% B% 0 90 10 1 90 10 25 0 100 30 0 100 - Flow Rate: 1.0 mL/min
- Wavelength: 280 nm
- Column Temperature: 30 °C
- Volume: 5 µL
- Sample: Disitamab Vedotin (ADC reference formulation)

2. DAR Analysis of Cysteine conjugated ADC
Chromatographic conditions:
- Column: Advanchrom Bio HIC-Butyl (4.6×100 mm, 5 µm)
- Mobile Phases:
A: 50 mM PB solution + 1.5 M ammonium sulfate, pH=7.0
B: 50 mM PB solution, pH=7.0 : isopropanol = 80:20 - Gradient Program:
Time (min) A% B% 0 100 0 20 0 100 25 0 100 - Flow Rate: 1.0 mL/min
- Wavelength: 280 nm
- Column Temperature: 30 °C
- Volume: 10 µL
- Sample: Cysteine conjugated ADC

3. Determination of Pembrolizumab
Chromatographic conditions:
- Column: Advanchrom Bio HIC-Butyl (4.6×100 mm, 5 µm)
- Mobile Phases:
A: 100 mM PB solution + 2.0 M ammonium sulfate, pH=7.0
B: 100 mM PB solution, pH=7.0 - Gradient Program:
Time (min) A% B% 0 90 10 10 90 10 15 0 100 20 0 100 - Flow Rate: 1.0 mL/min
- Wavelength: 280 nm
- Column Temperature: 30 °C
- Volume: 10 µL

4. Determination of Standard Protein Mixture
Chromatographic conditions:
- Column: Advanchrom Bio HIC-Butyl (4.6×100 mm, 5 µm)
- Mobile Phases:
A: 100 mM PB solution + 2.0 M ammonium sulfate, pH=7.0
B: 100 mM PB solution, pH=7.0 - Gradient Program:
Time (min) A% B% 0 90 10 10 90 10 15 0 100 20 0 100 - Flow Rate: 1.0 mL/min
- Wavelength: 280 nm
- Column Temperature: 30 °C
- Volume: 10 µL
- Samples: 1. Myoglobin; 2. RNase A; 3. Lysozyme; 4. α-Chymotrypsin

Welch Advanchrom Bio HIC-Butyl achieves good retention and separation among 4 standard proteins, with excellent batch-to-batch reproducibility, ensuring reliable analysis results.
Quality Assurance
Each Advanchrom Bio HIC-Butyl column is manufactured with a mature packing process and strict quality control to ensure superior separation performance and batch reproducibility.
Ordering Information
P/N | Product | Specification |
---|---|---|
0004-05039 | Advanchrom Bio HIC-Butyl | 5 µm, 4.6×100 mm |