Experimental report on measuring trans fatty acids using WM-CN100

Experimental report- measuring trans fatty acids by WM-CN100

Test Report

Chromatography conditions:

Column

WM-CN100 100m*0.25mm0.2μm07965-25011

Column Temperature

Speed ( ℃ /min)

Temperature (℃) 

Remain time(min)

 

140

5

1.8

220

20

 

Detector

FID 250 ℃

Injection port

250 ℃

Injection Volume

1μL

Flow Rate

1.3 ml/min

Carrier gas

Nitrogen

Split ratio

30-1

Hydrogen gas 

32mL/min

Air

200mL/min

Note

/

Chromatogram and Data: 

1) Mixture





Conclusion:

Using Welch WM-CN100 (100m*0.25mm, 0.2μm) 07965-25011 under this chromatographic condition, all peaks are normal, the peak order is consistent with the standard chromatogram, and the retention time is close to the reference, meeting the test requirements.