Open Photochemical Derivatives

#RTAreaHeightPlates(USP)Tailing factorCompound
16.5039636207305356651.082Bis-trioctylbenzene sulfonate
Total 96362073053   
ColumnWelch Xtimate C18 4.6x250mm,5μm  (P/N: 00101-21043)
Mobile PhaseMethanol: 0.5 mol/L Tris-HCl buffer (pH 8.9) = 85:15
DetectorFLD:excitation wavelength: 317nm, emission wavelength: 430nm
Temperature30℃
Flow Rate1.0ml/min
Injection Volume20ul
NotesDerivative samples are easily degraded