Ultisil Ion Exchange columns and media are silica based strong anion exchange (SAX – ammonium-functionalized silane) and strong cation exchange (SCX – covalently bonded aromatic sulfonic acid moiety) columns – with high resolution and high efficiency.
- Organic modifiers such as acetonitrile and methanol may be used with SAX and SCX columns, within organic/buffer solubility constraintsVery lipophilic
- Retention can be controlled by varying pH, ionic strength and organic modifier content.
Column Parameters
Columns | USP | pH range | Carbon Loading(%) | Surface Area(m2/g) | Endcapped | Max. Temp.(°C) |
Ultisil XB-SAX | L14 | 2.0-8.0 | 7.5(120Å), 1.5(300 Å) | 320(120Å), 90(300 Å) | No | 60(pH≤6.5), 40(pH>6.5) |
Ultisil XB-SCX | L9 | 2.0-8.0 | 12(120Å), 5(300Å) | 320(120Å), 90(300 Å) | No | 60(pH≤6.5), 40(pH>6.5) |
Applications
13 Heparin Disaccharides

Column | Ultisil XB-SAX, 3.0 ×250mm, 5µm |
Injection Volume | 10 µL |
Flow rate | 0.45mL/min |
Detector | 234 nm, 202 nm |
Temperature | 50°C |
Chondroitin Sulfate

Column | Ultisil XB-SAX, 4.6×250mm, 5µm |
Mobile Phase | A: water, adjust pH 3.5 with diluted HCl B: 2 mol/L NaCl, adjust pH 3.5 with diluted HCl |
Flow Rate | 1.0 mL/min |
Detector | 232 nm |
Temp. | Ambient |
Injection Volume | 20μL |
Mixed Standards: | Chondroitin disaccharide(B) 6- sulfated chondroitin disaccharide(C) 4- sulfated chondroitin disaccharide(A) |
Metformin HCL

Column | Ultisil XB-SCX, 4.6×250mm, 5µm |
Mobile Phase: | 1.7% NH4H2PO4( pH 3.0 adjusted by H3PO4) |
Flow rate | 1.0mL/min |
Detector | 218nm |
Temp. | Ambient |
Injection Volume | 10μL |
Samples In Order | Icyandiamide, melamine, metformin HCL |
Orazamide

Column | Ultisil XB-SCX, 4.6×250mm, 5µm |
Mobile Phase: | 1.0% NH4H2PO4( pH 3.0 adjusted by H3PO4) |
Flow rate | 1.0mL/min |
Detector | 215nm |
Temp. | 30°C |
Injection Volume | 10μL |
Samples In Order | Orotic acid, Pyridinepropanimidamide, 4-Amino-5-imidazolecarboxamide hydrochloride and AZO |
Catalog Downloads
Related Links
About us
Welch Materials is a multinational company that develops and manufactures chromatography consumables including HPLC column, Solid Phase Extraction (SPE) column, GC column, Prep column, Flash column and packing materials.
Our initial strength was our extensive experience on particle surface modification science and techniques. We are experts on bonding
chemistry and innovative packing materials for chromatography applications. Through the optimal utilization of our resources, we have developed many innovative LC and GC consumable products, in particular, our six series HPLC columns, Ultisil®, Welchrom®, Xtimate®, Topsil®, Boltimate® and Blossmate®.
Latest Posts
- 6 Easy Steps to Pump Maintenance – No More Waiting for Engineers Online
- What Happens to Ionic Compounds Inside a Chromatographic Column
- The Essentials of Recycling Rates
- New Product: Launches the Era of Diamond DLC Chromatography Columns
- Unveiling the Comprehensive Application of Derivative Techniques in Liquid Chromatography Analysis
- Choosing Chiral Separation Mode: Normal Phase or Reverse Phase?
- Solutions for Stucking and Leaking: Introducing the New Stainless Steel Connector
- Special Column for Synthetic Food Colorants – Restoring Food to its Original Appearance
- Selecting the Right Mechanical Validation Toolkit: A Comprehensive Guide
- Exploring Key Aspects of Dissolution Analysis Method Validation
- New UHPLC C18 Bonded Phase: The Preferred Choice for Routine Separation of Non-Retained or Strongly Retained Compounds
- Overcoming Challenges in SPE Experiments: Your Missing Piece for Success!