Solution of impurity A

1.1 Use leading peak suppressor

Gradient Program

Time(min)A%B%
01090
506040
ColumnWelch Ultisil XB-C18 4.6×250mm,5μm (P/N: 00201-31043) +Leading peak suppressor
Mobile PhaseA: Acetonitrile B: 50mmol/L ammonium acetate solution
Detector250nm
Temperature30℃
Flow Rate1ml/min
Injection Volume10ul
NotesNeed to use leading peak suppressor because of solvent effect, Or there will be double peak on impurity A

If you have any problem or require further information, please contact info@welchmat.com.