Solution of impurity A
1.1 Use leading peak suppressor
Gradient Program
Time(min) | A% | B% |
0 | 10 | 90 |
50 | 60 | 40 |
Column | Welch Ultisil XB-C18 4.6×250mm,5μm (P/N: 00201-31043) +Leading peak suppressor |
Mobile Phase | A: Acetonitrile B: 50mmol/L ammonium acetate solution |
Detector | 250nm |
Temperature | 30℃ |
Flow Rate | 1ml/min |
Injection Volume | 10ul |
Notes | Need to use leading peak suppressor because of solvent effect, Or there will be double peak on impurity A |
If you have any problem or require further information, please contact info@welchmat.com.