System suitability

Gradient Program

Time(min)A%B%
01000
150100
230100
241000
301000
ColumnXtimate® C18, 4.6×150 mm, 5 μm (P/N: 00101-21041)
Mobile PhasePhosphate buffer: 3.4 mL phosphate, 950 mL water, 4.3 mL 50% sodium hydroxide, add water to 1000mL and adjust pH to 7.10 with phosphoric acid or 50% sodium hydroxide solution. Mobile Phase A: Phosphate buffer/acetonitrile=98/2
Mobile phase B: Phosphate buffer/acetonitrile=78/22
Detection280 nm
Temperature35℃
Flow Rate1.3 mL/min
Injection Volume10 µL

If you have any problem or require further information, please contact info@welchmat.com.