Propofol system suitability solution HPLC- ELSD in reference literature

Peak 2: LPE   Peak 3: Propofol   Peak 4: LPC

ColumnUltisil® Diol 4.6×250 mm, 5 μm    P/N: 00206-31043
DetectionELSD: drift tube 70 ℃, carrier gas: 2.5L/min
Temperature40 ℃
Flow Rate1.0 ml/min
Injection Volume20 µl

Mobile Phase

A: Methanol-Water- Glacial Acetic Acid- Triethylamine=85:15:0.5:0.05

B: N-Hexane- Isopropanol-Mobile Phase A=20:48:32

TimeA%B%
0595
102278
229010
23595
27595