System Suitability
Column | Ultisil Polar RP, 5μm, 4.6×250mm (P/N 00215-31043) |
Mobile Phase | Acetonitrile : 0.1% trichloroacetic acid (adjust pH to 2.1 with phosphoric acid)=21.5:78.5 |
Detection | 349nm |
Temperature | 30℃ |
Flow Rate | 1.0ml/min |
Injection Volume | 20ul |