Related substance of test sample
Gradient Program
Time(min) | A% | B% |
0 | 70 | 30 |
20 | 0 | 100 |
35 | 0 | 100 |
50 | 70 | 30 |
Column | Welch Ultisil XB-ALK 4.6×250mm,5μm (P/N: 00253-31043) |
Mobile Phase | A: 0.5 mol/L potassium dihydrogen phosphate (adjust pH to 3.5 with phosphoric acid ) – methanol – Water = 10:30:60 B: 0.5 mol/L potassium dihydrogen phosphate (adjust pH to 3.5 with phosphoric acid ) – methanol – Water = 10:50:40 |
Detector | 225nm |
Temperature | 30℃ |
Flow Rate | 1ml/min |
Injection Volume | 20ul |
Notes | Sensitive to injection volume and dissolved sample solvent |
If you have any problem or require further information, please contact info@welchmat.com.