Gradient Condition
Time(min) | A% | B% |
0 | 45 | 55 |
20 | 45 | 55 |
40 | 10 | 90 |
Column | Welch Ultisil® PFP, 3 um, 4.6×150 mm |
Mobile Phase | Mobile Phase: A Mix 0.5 mL trifluoroacetic acid with 500 ml water of and filter it B: Pure methanol |
Wavelength | 215 nm |
Temperature | 40℃ |
Flow Rate | 1.0 mL/min |
Injection Volume | 10 μL |