Mixed standards(2ug/mL)

CompoundsRtAreaHeightPlates(USP)TailingR(USP)
7-ACP1.487486431249025561.173 
Tert-butyl ester1.688471731132028261.1171.650
Ceftazidime2.552548221120650511.0696.383
D-T-ACA2.851504891043964701.0712.093
7-ACA3.39142551791179351.0433.672
Cefuroxime4.64246049684397090.9677.343
Cefuroxime deacetylated5.133479786641106400.9752.533
7-ACT9.791733725581135431.01217.395
ceftriaxone12.227655273936127000.8776.325
T-ACA16.64917325821148241.0309.016
Cefodizime20.724442241688146450.9516.615
ColumnXtimate® C18, 4.6×250 mm, 5 μm (P/N: 00101-21043)
Mobile PhasePhosphate buffer/ acetonitrile =500/500 (V/V)
Detection254nm, 215nm
Temperature30℃
Flow Rate1.5 mL/min
Injection Volume20 µL
Notes/