System adaptability solution
Column | Welch Ultisil LP-CN 4.6×250mm,5μm (P/N: 00247-31043) |
Mobile Phase | Acetonitrile: Buffer [take 1.44 g sodium lauryl sulfate and 0.75 g tetrabutylammonium bromide, add water,dissolve and dilute to 1000ml, adjust the pH to 3.0 with a 0.05mol/L sulfuric acid solution]=22:78 |
Detector | 230 nm |
Temperature | 35℃ |
Flow Rate | 1.0 ml/min |
Injection Volume | 20µl |
Notes | Need long balance time because of ion pair reagent in mobile phase |