Test Solution

ColumnUltisil® XB-C18(4.6×250mm,5um);
Detection254nm
Temperature20℃
Flow Rate1.0ml/min
Injection Volume10u
Note/

Mobile Phase

Time(minMobile phase A (%)Mobile phase B (%)
07030
57030
154456
354456
35.57030
607030