ColumnXtimate® C18(4.6*250mm,5μm)
Mobile PhaseSee the table below for details
Detection200 nm/280 nm
Temperature30℃
Flow Rate1.0 ml/min
Injection Volume50 μL
Note1. If 200 nm wavelength is selected, use higher quality chromatography grade
methanol.
2. After comparative testing, the response of VB2 at 280nm is greater than 200nm,
and the other five substances have higher response at 200nm wavelength.

Mobile Phase: (A: 30 mmol/L potassium phosphorus dihydrogen (adjust pH to 5.0) B: methanol)

Time (min)Mobile phase A(%)Mobile phase B(%)
0955
157723
205050
255050
25.01955
35955