Column | Ultisil® LP- C18(4.6×250mm,5μm) |
Mobile Phase | See the table below for details |
Detection | 220nm |
Temperature | 20℃ |
Flow Rate | 1.0 ml/min |
Injection Volume | 10 μL |
Mobile Phase:
Mobile phase A: Weigh 3.48g of dipotassium hydrogen phosphate, dissolve it in 1000ml of Wahaha water, adjust the pH to 7.0 with phosphoric acid, mix well, filter with suction, and degas by ultrasonic Mobile phase B: Acetonitrile.
Time (min) | Mobile phase A(%) | Mobile phase B(%) |
0 | 90 | 10 |
6 | 84 | 16 |
10 | 78 | 22 |
40 | 75 | 25 |
50 | 55 | 45 |
51 | 90 | 10 |
65 | 90 | 10 |