ColumnUltisil® LP- C18(4.6×250mm,5μm)
Detection220nm
Temperature20℃
Flow Rate1.0ml/min
Injection Volume10ul
Note/

Mobile Phase

Mobile phase A: weigh 3.48g of dipotassium hydrogen phosphate, dissolve it in 1000ml water, adjust the pH to 7.0 with phosphoric acid, mix well, filter by suction, and degass by ultrasound;

Mobile phase B: acetonitrile

Time(min)Mobile phase A (%)Acetonitrile (%)
09010
68416
107822
407525
505545
519010
659010