Triethylamine is a conventional solvent that is applied in different fields. As tailing phenomenon often occurs during the detection of amine substances, we prepared the test of triethylamine of the same concentration by different polar columns as below.

Chromatographic Condition

Column temperature

Flow rate(℃/min)Temp(℃)Rt(min)
 _4010
202003

Injection inlet200℃
DetectorFID 250℃
Carrier gasNitrogen
Column flow rate1.0ml/min
Injection method(10-1)

Chromatogram and Data

Column: WM-FFAP  30m*0.25mm*0.25μm

No.CompoundRtHeightAreaArea%
1Triethylamine3.3474842.3711342.62100.0000
  _Total  _4842.3711342.62100.0000

No.CompoundRtCapacity factor PlatesResolutionTailing factor
1Triethylamine3.3470.0000613930.0001.769

Column: WM-FFAP  30m*0.25mm*0.25μm

No.CompoundRtHeightAreaArea%
1Triethylamine13.54317.95530.58100.0000
  _Total  _17.95530.58100.0000

No.CompoundRtCapacity factor PlatesResolutionTailing factor
1Triethylamine13.5430.000052359.7785.043

Column: WM-FFAP  30m*0.25mm*0.25μm

No.CompoundRtHeightAreaArea%
1Triethylamine8.7111822.9611692.22100.0000
  _Total  _1822.9611692.22100.0000

No.CompoundRtCapacity factor PlatesResolutionTailing factor
1Triethylamine8.7110.0000428690.0000.942

Conclusion

WM 5-Amine column (30m*0.32mm*1.0μm) had good peak shape and column efficiency in this detection of triethylamine. As this kind of substance may also have residue in the system, it is recommended to clean or replace instrument components after detection.