Blank solvent

Column | Ultisil® Alk C18( 4.6×300mm,5μm) |
Detection | 205nm |
Temperature | 30℃ |
Flow Rate | 1.0ml/min |
Injection Volume | 30ul |
Note | / |
Mobile Phase
Mobile phase A: Take 23 g of phosphoric acid to add 1000 ml of water to prepare a 23 g / L phosphoric acid solution. Take 5 ml of chromatographic acetonitrile, 125ml23 g / L phosphoric acid solution and 870ml water mix, shake well, ultrasonic degassing.
Mobile phase B: Take 23 g of phosphoric acid to add 1000 ml of water to prepare a 23 g / L phosphoric acid solution. Take 475 ml of chromatographic acetonitrile, 125ml23 g / L phosphoric acid solution and 400ml water mixed, shake well, ultrasonic degassing.
Time(min) | Mobile phase A (%) | Mobile phase B (%) |
0 | 100 | 0 |
6 | 100 | 0 |
50 | 0 | 100 |
60 | 0 | 100 |
70 | 100 | 0 |
80 | 100 | 0 |
If you have any problem or require further information, please contact info@welchmat.com
0 Comments