ColumnWM-FFAP (30m×0.25mm, 0.25μm) (P/N: 03911-22001)
TemperatureSee the table below for details
Injector260℃
DetectorFID 260℃
Carrier gasNitrogen
Flow rate1.0mL/min
Split ratio20-1
Injection volume1μL
Hydroge30mL/min
Air300mL/min

Temperature

Rate (℃/min)Temperature (℃)Retention time (min)
1001
522020