Blank solution

Column | Xtimate® C18( 4.6×250mm,5μm) |
Detection | 205nm |
Temperature | 30℃ |
Flow Rate | 1.0ml/min |
Injection Volume | 20ul |
Note | / |
Mobile Phase
Mobile phase A: weigh 1.36g of potassium dihydrogen phosphate, dissolve it in 1000ml of water, add 12ml of 25% tetrabutylammonium hydroxide aqueous solution, adjust the pH value to 7.0 with phosphoric acid, mix well, filter by suction and Degas by ultrasonic. Mobile phase B: take the color spectrum of pure acetonitrile and degassing by ultrasonic.
Time(min) | Mobile phase A (%) | Acetonitrile (%) |
0 | 93 | 7 |
25 | 93 | 7 |
40 | 40 | 60 |
50 | 40 | 60 |
55 | 93 | 7 |
70 | 93 | 7 |
If you have any problem or require further information, please contact info@welchmat.com.