ColumnUltisil® XB-C18(
4.6×300mm,5μm)
Detection240nm
Temperature25℃
Flow Rate1.0ml/min
Injection Volume20ul
Note/

Mobile Phase
Mobile phase A: Weigh 2.72g potassium dihydrogen phosphate, dissolve in 1000ml water, adjust with phosphoric acid pH value to 3.0, mix well and suction filter.

Time(min)Mobile phase A (%)Acetonitrile (%)
07030
156040
255050
404060
503070
553070
587030
707030