Sample Solution
Column | Ultisil® Diol 4.0×125 mm, 5 μm P/N: 00206-31031 |
Detection | evaporative light-scattering detector: drift tube 60 ℃, carrier gas: 2.0L/min |
Temperature | 55 ℃ |
Flow Rate | 1.0 ml/min |
Injection Volume | 20 µl |
Mobile Phase
A: n-hexane:200mL, isopropanol 42.5mL, glacial acetic acid 3.8mL, triethylamine 0.2 mL
B: isopropanol 170 mL, glacial acetic acid 3mL, water 28mL, triethylamine 0.16 mL
Time | A% | B% |
0 | 95 | 5 |
5 | 80 | 20 |
8.5 | 60 | 40 |
14 | 55 | 45 |
15 | 0 | 100 |
17.5 | 0 | 100 |
17.6 | 95 | 5 |
40 | 95 | 5 |