Reference solution

ColumnBoltimate EXT-C18 (150 x 4.6)mm 2.7µ
RT1.7 min
Temperature30°C
Flow Rate1.0ml/min
Injection Volume10µL
Wavelength210 nm
Run Time6 min

Results

Tailing factor1.24 (Limit : NMT 2)
Theoretical Plates1.41 (Limit : NMT 1.5)
% RSD0.1% (six replicate standard solution) (Limit : NMT 1.0%)
Maximum Back Pressure270 Bar