Test Sample

ColumnXtimate® C18, 4.6×100 mm, 3μm
Detection215nm
Temperature35℃
Flow Rate1.0 mL/min
Injection Volume8 µL

Mobile Phase

Time(min)Dipotassium hydrogen phosphate (pH8.9)Acetonitrile
05545
45545
73070
143070
185545