Blank excipient solution

ColumnUltisil® UHPLC Polar RP(
2.1×100mm,1.8μm)
Mobile PhaseMobile phase: take 11.16g potassium dihydrogen phosphate and 7.2g
potassium dihydrogen phosphate trihydrate, add 1000ml Wahaha
water, mix well and filter; Take 880ml of the above solution and
120ml of chromatographic grade methanol, adjust the pH to 3.0 with
1mol / L phosphoric acid, and degass with ultrasound.

Phosphate buffer / methanol = 88 / 12
Detection280nm
Temperature35℃
Flow Rate0.15ml/min
Injection Volume2ul
Note/