Reference solution

Column | Boltimate C18 (100 x 4.6)mm 2.7µ |
RT | 4.88min |
Temperature | Ambient |
Flow Rate | 1.0ml/min |
Injection Volume | 10ul |
Wavelength | 285 nm |
Run Time | 15min |
Results
Tailing factor | 1.24 (Limit : NMT 2) |
Theoretical Plates | 10365 (Limit : NLT 4000) |
Maximum Back Pressure | 172 Bar |
If you have any problem or require further information, please contact info@welchmat.com.