Column | WEL-1301 (50m×0.53mm,1.0μm)(P/N:01905-53003) |
Temperature | See the table below for details |
Injector | 250℃ |
Detector | FID 250℃ |
Carrier gas | Nitrogen |
Flow rate | 4.0mL/min |
Split ratio | No split |
Injection method | Headspace injection |
Injection volume | 1mL |
Equilibrium temperature | 80℃ |
Line temperature | 100℃ |
Valve box temperature | 100℃ |
Equilibrium time | 20min |
Acquisition time | 60min |
Hydrogen | 30mL/min |
Air | 300mL/min |
Temperature:
Rate (℃/min) | Temperature (℃) | Retention time (min) |
50 | 4 | |
30 | 200 | 40 |