ColumnWEL-1301 (50m×0.53mm,1.0μm)(P/N:01905-53003)
TemperatureSee the table below for details
Injector250℃
DetectorFID 250℃
Carrier gasNitrogen
Flow rate4.0mL/min
Split ratioNo split
Injection methodHeadspace injection
Injection volume1mL
Equilibrium
temperature
80℃
Line temperature100℃
Valve box
temperature
100℃
Equilibrium time20min
Acquisition time60min
Hydrogen30mL/min
Air300mL/min

Temperature:

Rate (℃/min)Temperature (℃)Retention time (min)
504
3020040