Column | Ultisil® ODS-3(4.6×250mm,5μm) |
Detection | 254nm |
Temperature | 35℃ |
Flow Rate | 1.0ml/min |
Injection Volume | 10ul |
Note | / |
Mobile Phase
0.01mol/l ammonium formate: take 0.63g ammonium formate, dissolve it in 1000ml of water, and use a Adjust the pH of acid to 3.0, shake well and degass with ultrasonic. Acetonitrile: take chromatographic grade acetonitrile to obtain
Time(min) | 0.01mol/l ammonium formate | Time (min): 0.01mol/l ammonium formate acetonitrile |
0 | 96 | 4 |
8 | 96 | 4 |
18 | 90 | 4 |
23 | 90 | 10 |
53 | 65 | 35 |
58 | 65 | 35 |
59 | 96 | 4 |
70 | 96 | 4 |