Blank solution diagram

Column | Ultisil® PFP (4.6×250mm,5μm) |
Detection | 258nm |
Temperature | 30℃ |
Flow Rate | 1.0ml/min |
Injection Volume | 20ul |
Note | / |
Mobile Phase
Phosphate buffer solution: take 0.02mol/l potassium dihydrogen phosphate and 0.02mol/l potassium dihydrogen phosphate, mix them 1 / 1, adjust the pH to 6.5 with 20% phosphoric acid solution, mix them, and filter them.
Time(min) | Phosphate buffer (%) | Acetonitrile (%) |
0 | 100 | 0 |
10 | 100 | 0 |
30 | 70 | 30 |
35 | 70 | 30 |
36 | 100 | 30 |
45 | 100 | 0 |
If you have any problem or require further information, please contact info@welchmat.com.