System suitability solution

Column | Ultisil® XB-C18( 4.6×250mm,5μm) |
Mobile Phase | Ammonium dihydrogen phosphate solution (ph4.4): 20.7g ammonium dihydrogen phosphate, add 1800ml water to dissolve it, adjust the pH value to 4.4 with phosphoric acid, mix well, and filter. Ammonium dihydrogen phosphate solution (ph4.4) / acetonitrile = 90 / 10 |
Detection | 280nm |
Temperature | 30℃ |
Flow Rate | 1.0ml/min |
Injection Volume | 20ul |
Note | / |
If you have any problem or require further information, please contact info@welchmat