Application Range
It is suitable for the determination of acetylsalicylic acid in facial paste, cream and lotion cosmetics for whitening and moisturizing purposes (the substrate used here is lotion).
Reference standard: Detection of acetylsalicylic acid in imported cosmetics SN/T 2290-2009
Extraction Steps
Weigh 1g sample accurately(0.01g) in 15 ml centrifuge tube with plug. Then add 10 ml methanol and mix well. Extract by ultrasound for 10 min, centrifuge at 8000r/min for 5 min, take the supernatant to filter through 0.22um filter membrane to be tested.
Chromatographic Condition
Column: Ultisil@XB-C18, 4.6*250mm, 5um
Mobile phase: acetonitrile: tetrahydrofuran: acetic acid: water =20:5:70
Flow rate:1.0mL/min
Column temp: 30℃
Injection volume: 10μL
Wavelength: 303 nm
Chromatogram and Data
Acetylsalicylic acid 5mg/L

Peak | Rt | Area | Height | Area | Tailing factor | Plates |
_ | min | mAU*min | mAU | % | EP | (USP) |
Acetylsalicylic Acid | 14.411 | 0.9689 | 3.48 | 100 | 0.98 | 16684 |
Blank cosmetic water sample

Blank cosmetic water sample spiked at 50mg/kg

Spike recovery rate table
Name | Spike Level | Average Recovery Rate | RSD |
_ | mg/kg | % | %(n=4) |
Acetylsalicylic Acid | 50 | 101 | 1.62 |
Ordering Information
P/N | Name | Specification |
00837-05002 | 15mL centrifuge tube with screw mouth and apical bottom | Disposable centrifuge tube, flat cap,conical bottom, RCF12000xg,15 mL, 50/ pk without sterilization |
00821-32291 | Caps and septa | Pre-slit red PTFE/white Silicone septa, 9mm blue short screw-thread polypropylene cap, 6mm centre hole, 100/pk |
00821-40927 | Vials | 2ml wide opening short screw-thread vial with write-on spot, clear, 11.6*32mm , 100/pk |
00201-31041 | HPLC Column | Ultisil® XB-C18,4.6×250mm,5μm |