Overview

The organic impurity test of Azithromycin API is given in USP monograph. In this method, it was mentioned to use ‘4.6 mm x 25 cm; 5 µm packing L1’ HPLC column. From USP column database, we obtained that USP recommended to use Waters XTerra MS C18, 4.6 mm X 25 cm, 5µm HPLC column. The method equivalency study has been carried out for related substances of Azithromycin API USP to establish equivalency between Welch Xtimate® HPLC column andWaters XTerra HPLC Column. This report presents and discusses the results obtained in the equivalency study.

Columns

  1. Welch Xtimate® C18, (250 x 4.6)mm, 5µm
  2. Waters X-terra MS C18, (250 x 4.6), 5µm

Chromatographic system

Column1. Welch Xtimate C18 (250 mm x 4.6 mm), 5 µm
2. Waters X-terra MS C18, (250 mm x 4.6 mm), 5µm
Wavelength210 nm
Flow rate1.0 mL/minute
Injection volume50 µL
Column temperature60°C
Auto Sampler temperature25°C

Results

Column-1 : Welch Xtimate C18 (250 mm x 4.6 mm), 5 µm

Column-2 : Waters X-terra MS C18, (250 mm x 4.6 mm), 5µm

Conclusion

From the results of the equivalency study, the equivalency between the Welch Xtimate® HPLC column and Waters XTerra HPLC Column is proved. Hence a Welch Xtimate® HPLC Column can be used for related substances by HPLC of Azithromycin API as per USP.

Chromatograph of Blank

Fig 1: Welch Xtimate C18 (250 mm x 4.6 mm), 5 µm

Fig 2:Waters X-terra MS C18, (250 mm x 4.6 mm), 5µm

Chromatography of system suitability Solution

Fig 3:Welch Xtimate C18 (250 mm x 4.6 mm), 5 µm

Fig 4: Waters X-terra MS C18, (250 mm x 4.6 mm), 5µm

Chromatograph of Standard Solution

Fig 5: Welch Xtimate C18 (250 mm x 4.6 mm), 5 µm

Fig 6: Waters X-terra MS C18, (250 mm x 4.6 mm), 5µm

Chromatograph of Sample Solution

Fig 7:Welch Xtimate C18 (250 mm x 4.6 mm), 5 µm

Fig 8: Waters X-terra MS C18, (250 mm x 4.6 mm), 5µm

If you have any problem or require further information, please contact info@welchmat.com.