Mixed standard solution

Gradient Program

Time(min)A%B%
07030
324060
344060
367525
457525
ColumnWelch Ultisil Plus C18 4.6×250mm,5μm (P/N: 00260-31043)
Mobile PhaseMobile phase A: 4.76g/L potassium dihydrogen phosphate (pH4.4) Mobile phase B: Acetonitrile
Detector205nm
Temperature40℃
Flow Rate1.1 ml/min
Injection Volume10µl