Clarithromycin Capsules on Ultisil XB-C18, AQ-C18 and Xtimate C18

Ultisil XB-C18, 4.6×250mm, 5μm Content Determination

Column Ultisil XB-C18, 4.6×250mm, 5μm (P/N 00201-31043) Ultisil AQ-C18, 4.6×250mm, 5μm (P/N 00207-31043) Xtimate C18, 4.6×250mm, 5μm (P/N 00101-21043)
Mobile Phase Phosphate Buffer :Acetonitrile = 60:40
Detection 210nm
Temperature 45°C
Flow Rate 1.0ml/min
Injection Volume 20µl
Notes Use PTFE filter or centrifuge to filtrate sample solution