Ghost-Buster Column
  • Overview
  • Parameters
  • Order
  • Application
  • Materials
  • Instructions

In HPLC analysis, especially in gradient elution, some unexpected peaks, which are usually called Ghost peaks, may appear in a chromatogram.

Ghost peaks my appear for many reasons, including residual in injection valve, water pollution impurities and unknown components in sample. Generally, they are caused by impurities from mobile phase.

Welch Ghost-Buster Column can solve this problem easily!

Welch Ghost-Buster column can efficiently adsorb and remove the impurities from mobile phase to eliminate their interference to the target peaks. Installed between the gradient mixer and the sampler, Ghost-Buster column can trap most ghost peaks before HPLC analysis.

Applications:

1.

Column:UltisilTM XB-C18, 4.6×250 mm, 5 µm

Flow Rate: 1.0 ml/min

Injection Volume: 10 µl

Detection: 210 nm

Temperature: 40 ℃

Sample Preparation:Ultra-pure water

Mobile Phase: A.Ultra-pure water; B. Acetonitrile

Gradient:

Time(min)

A%

B%

0

90

10

20

10

90

30

10

90

30.1

90

10

38

90

10


a: without Ghost-Buster column; b: with Ghost-Buster column (4.6×50 mm)


2.

Column:UltisilTM XB-C18, 4.6×250 mm, 5 µm

Flow Rate: 1.0 ml/min

Injection Volume: 20 µl

Detection: 280 nm

Temperature: 40 ℃

Sample Preparation:Ultra-pure water

Mobile Phase: A.1.54 g/l Ammonium ethanoate solution (adjust pH to 4.0 with GAA); B. Acetonitrile

Gradient:

Time(min)

A%

B%

0

98

2

10

80

20

20

40

60

35

20

80

40

20

80

41 

 98

 2

 50

 98

 2


a: without Ghost-Buster column; b: with Ghost-Buster column (4.6×50 mm)


3.

Column:UltisilTM XB-C18, 4.6×150 mm, 3 µm

Flow Rate: 1.0 ml/min

Injection Volume: 20 µl

Detection: 254 nm

Temperature: 30 ℃

Sample Preparation:Ultra-pure water

Mobile Phase: A.20 mmol/l DAP solution (adjust pH to 6.0 with phosphoric acid); B. Acetonitrile

Gradient:


Time(min)

A%

B%

0

85

15

30

45

55

30.1

85

15

40

85

15




a: without Ghost-Buster column; b: with Ghost-Buster column (4.6×50 mm)

productName OrderNo Package Price Instock
Ghost-Buster HP Column 06100-31025 2.1×50 mm Inquiry Inquir Order
Ghost-Buster HP Column 06100-31021 2.1×33 mm Inquiry Inquir Order
Ghost-Buster Column 06100-31001 7.8×50 mm Inquiry Inquir Order
Ghost-Buster Column 06100-31000 4.6×50 mm Inquiry Inquir Order