Ultisil XB-C4-Peptide Side Chain Analysis Method Validation Report.

Test Report

Chromatography conditions:

Column 1

Xtimate XB C44.6*250mm*5μm300A(SN:60220601601LN:2803.15

Column 2

Xtimate XB C4 4.6*250mm*5μm300A(SN:60230403506LN:2803.16)

Column 3

Xtimate XB C4 4.6*250mm*5μm300A(SN:60230403507LN:2803.16)

Mobile Phase

Phase A: 0.01% formic acid aqueous solution

Phase B: chromatographically pure acetonitrile

Column Temperature

40℃

Detector

UV

Injection Volume

20μL

Flow Rate

1.0 ml/min

Detect wavelength

214 nm

Elution procedure

Time (min)

Ammonium sulfate buffer: acetonitrile = 90:10

Ammonium sulfate buffer: acetonitrile = 50:50 (%)

0

90

10

35

0

100

50

0

100

51

90

10

60

90

10


Note

After the analysis of the above batch of chromatographic columns, use the initial proportion of mobile phase 0.2ml/min flow rate to flush for 24 hours and then analyze again.

Chromatogram and Data: 

1) Chromatographic column 1 (SN: 60220601601) analyzes the test solution chromatogram:

 Serial number 

Peak name 

Retention time (min)

Peak Area(mAU*min)

S/N

Relative peak area(%)

Resolution(EP)

Number of plates(EP)

Asymmetry(EP)

1

Main materials

27.242

70.234

196.8

98.05

19.02

397233

1.21

2

Abnormal peak

33.765

1.399

1.0

1.95

N.a.

68855

0.81

Sum


 

71.633

 

100.00

19.02

 

 


2) Chromatographic column 2 (SN: 60230403506) and chromatographic column 3 (SN: 60230403507) analyze the test solution chromatogram:

Serial number

Retention time (min)

Peak Area(mAU*min)

S/N

Relative peak area(%)

Resolution(EP)

Number of plates(EP)

Asymmetry(EP)

1

27.555

73.030

207.6

98.49

10.48

353181

1.08

2

29.838

1.121

2.9

1.51

N.a.

225042

0.75

Sum


74.151

 

100.00

10.48

 

 

  • 3) Analyze the test solution chromatogram after low-speed flushing of the mobile phase with the above initial ratio of the chromatographic column.

 

Serial number

Peak name 

Retention time (min)

Peak Area(mAU*min)

S/N

Relative peak area(%)

Resolution

(EP)

Number of plates(EP)

Asymmetry(EP)

1

Main materials

27.158

71.999

203.2

100.00

N.a.

392779

1.15

Sum



71.999

 

100.00

0.00

 

 

Conclusion:

Using different batches of Welch hpcl column Xtimate XB-C4 (4.6×250mm, 5μm300A), under this chromatographic condition, analyze the solution of the test sample. An abnormal chromatographic peak appears at a retention time of 29-34 minutes. After flushing with the initial proportion of mobile phase at a flow rate of 0.2ml/min for 24 hours, reanalyze the solution of the test sample. No abnormal chromatographic peaks appear during the corresponding time period.