New experimental report on thymus assay using Ultisil LP-C18

Test Report

Chromatography conditions:

Column 

Welch Ultisil®LP-C18,(4.6×250mm,5μm

Mobile Phase

Ammonium sulfate-acetonitrile

Column Temperature

50℃

Detector

UV

Injection Volume

20μL

Flow Rate

1.0 ml/min

Detect wavelength

210 nm

Elution procedure

Time (min)

Ammonium sulfate buffer: acetonitrile = 90:10

Ammonium sulfate buffer: acetonitrile = 50:50 (%)

0

88

50

45

80

20

50

50

50

51

88

12

60

88

12


Note

/

Chromatogram and Data: 

1) Blank

2) System adaptation:

Retention time (min)

Peak Area

Peak height

Resolution(USP)

Tailing factor

Number of plates(EP)

26.816

607.38

28.75

 

1.09833

36707.33671

27.692

6615.32

299.36

1.52562

1.18793

35311.8687

28.923

9.62

0.44

2.14121

1.37611

42759.16405

30.219

30.96

0.96

1.83447

0.79124

19856.50640

Sum

7263.28

 

 

 

 


Conclusion:

Under these HPLC conditions,Using Welch Ultisil® LP-C18, (4.6×250mm, 5μm), the retention time of the main peak is about 30min, and the impurity separation is greater than 1.5. The verification results of the latest batch of two chromatographic columns are repeatable and meet the test requirements.